Calibration Certificate
This calibration certificate gives an overview of the results obtained during the calibration.
Fluid no:
The fluid position in the instrument where the calibrated fluid is available.
Calibration by comparison:
The device under test has been compared with a calibration standard.
Calibration date:
Date of the calibration.
The traceability of a calibration
VSL:
VSL is the National Metrology Institute of The Netherlands. VSL maintains and develops the Dutch measurement standards.
Bronkhorst Calibration Laboratory (BCL):
BCL maintains the measurement standards of Bronkhorst, which are calibrated by VSL.
Bronkhorst Production:
The calibration standards used in Bronkhorst's production process are calibrated by BCL.
The calibrated instrument.
Type:
Is it a Flow meter or flow controller.
Serial number:
Serial number or order number.
Model number:
Type of instrument (customer model or Bronkhorst model).
Rated accuracy*:
The accuracy of this instrument specified under calibration conditions.
Calibration Standard:
The calibration standard wich is used during the calibration (example: Piston Prover, Rotor Meter, Turbine Meter, Reference Meter, etc.)
Type:
The calibration device wich is used during the calibration, in this example a piston prover.
Serial number:
Serial number of the calibration device.
Certificate no.:
The certificate of the calibration device, which certifies the validity of the calibration standard
Uncertainty:
The reported uncertainty of measurement is based on the standard uncertainty multiplied by a coverage factor k=2.
Customer conditions
Fluid and conditions as indicated by customer.
Fluid:
The fluid (specified by customer)
Flow:
The flow range (specified by customer)
Pressure:
The pressure (specified by customer)
Temperature:
The temperature (specified by customer).
Calibration conditions
The calibration conditions shows under which conditions the instrument is calibrated.
Fluid:
The calibration fluid.
Flow:
The calibration flow range. ( Equivalent flow: calibrated flow equivalent to customer flow)
Pressure:
The calibration pressure.
Temperature:
The temperature during the calibration.
Atm. Pressure:
The atmospheric pressure during the calibration.
Calibration and Conversion Results
Point:
Calibration point
Calibrated flow AiR:
Fluid and flow point used during calibration.
Conversion factor:
A calculation from the reference gas flow rate to customer flow rate.
Customer flow** AiR:
The flow range (ordered by customer) converted from the calibrated flow (see:**)
Output signal:
The output signal flow range (ordered by customer) corresponding to the customer flow.
The analog signal is calculated based on the digital signal.
Calculation Customer Flow:
Calibrated flow • Conversion Factor
Measurement uncertainty of the calibrated flow AiR:
Total measurement uncertainty of the calibrated flow, including the calibration standard and the measurement setup.
Measured deviation:
Maximum measured deviation between the reference flow and the flow indicated by the calibrated instrument over the entire calibrated flow range.
Calibrator
The person who did the calibration.
Date:
The date when the certificate was checked.
QC:
Quality check by authorized person